28
PCD-BA-e-1211
8.3.3 Selection of adjustment weight
In the model series KERN PCD, the adjustment weight can be
selected from pre-set nominal values (refer also to following table,
factory setting with grey background). In order to achieve high-
quality weighing results in the sense of the measuring technology,
it is recommended to select the nominal value as high as
possible. The non delivered adjustment weights can be
purchased from KERN as option.
In weighing mode keep the
key pressed until
[Unit]
appears.
Press the
MODE
key several times until „
CAL
“ is displayed.
Acknowledge using
SET
key, the current setting is displayed.
Select the desired settings by pressing the
MODE
key.
Use the
SET
key to confirm selection.
Possible Adjustment Points:
PCD 250-3
PCD 400-3
PCD 2500-2
PCD 4000-2
PCD 10K0.1
-
100 g
-
-
-
50 g
200 g
500 g
0.5 kg
-
100 g
250 g
1000 g
1 kg
2 kg
150 g
300 g
1500 g
2 kg
5 kg
200 g
350 g
2000 g
3 kg
7 kg
250 g
400 g
2500 g
4 kg
10 kg