PCD-BA-e-1211
35
9
Data output RS 232 C
9.1 Technical data
8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bit, no parity bit
Baud rate selectable at 1200, 2400, 4800,
9600
and 19200 Baud
Miniature plug-in necessary (9 pole D-Sub)
For operation with interface faultless operation is only ensured with the correct
KERN
– interface cable (max. 2m)
9.2 Pin allocation of balance output bushing:
Front view:
5
1
9 6
Pin 2:
Transmit data
Pin 3:
Receive data
Pin 5:
Signal ground