38
PCD-BA-e-1211
Symbols
9.4
Output on bar code printer
The data transfer mode has to be set on
„BA Pr“
(chapter 8.4.1).
As bar code printer the Zebra printer model LP2824 is provided.
Take into account that the output format of the balance is fixedly defined and cannot
be changed.
The printer format is stored in the printer, i.e. in case of a failure the printer cannot be
changed with a new one from factory, previously it is necessary that KERN installs
the respective software.
The Zebra printer and the balance must be connected to the delivered interface
cable when they are switched off.
After switching-on both appliances, and after reaching the status ready-for-operation,
a label will be printed out when pressing the
button.
M
Blank or M
S
Space character or negative sign (-)
N
1
… N
10
10 numeric ASCII codes for weight values including decimal places or
blanks
U
1
… U
3
3 ASCII codes for weighing unit pcs. / % / or blank
B
Blank
E, o, r
ASCII code or “E, o, r”
CR
Carriage Return
LF
(Line Feed)