YBI-01A-BA-e-1620 

 

GENERAL 

The ionizer has conductive peaks supplied by high 
voltage which, based on corona discharge, generate 
positively and negatively charged ions in the immediate 
vicinity. These are attracted by the electrostatic charge of 
goods to be weighed and, thus, neutralise the interfering 
electrostatic charge. This also does away with the forces 
falsifying the weighing (such as falsified weighing result, 
weighing value drifted).  

 

2  Basic Safety Precautions 

 

WARNING

 

 

The use of ionizer is only intended in combination with 
electronic weighing balances.  Do not use for any other 
purposes. 

 

Never operate the ionizer in explosive environment. The 
serial version is not explosion protected. 

 

Protect the ionizer against high air humidity / 
temperature, steams and dust;  

Take care to select a location free of water and oil 

Do not expose the ionizer to strong humidity for 
extended periods. Non-permitted condensation 
(condensation of air humidity on the appliance) may 
occur if a cold appliance is taken to a considerably 
warmer environment. In this case, acclimatize the 
disconnected ionizer for ca. 2 hours at room 
temperature. 

 

Whilst the ionizer is switched on, do not touch the ion 
source; see sticker on the left side.  

 

In case of smoke development, smell of fire, strong 
heating-up of the ionizer or when the red LED starts 
glowing, turn off the ionizer immediately and disconnect 
it from the mains.  

 

If water or other foreign matter enters the ionizer, turn 
off the master switch immediately and disconnect it from 
the mains.  

 

 

Summary of Contents for YBI-01A

Page 1: ...n GmbH Ziegelei 1 D 72336 Balingen email info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating Manual Ionizer KERN YBI 01A Version 2 0 10 2016 GB YBI 01 BA e 1620 ...

Page 2: ... PRECAUTIONS 3 3 TECHNICAL DATA 6 4 APPLIANCE OVERVIEW 7 5 UNPACKING SETUP AND COMMISSIONING 8 Unpacking 8 Placing 8 Mains connection 9 6 COMMISSIONING 9 Turning on the ionizer 9 Turning on the fan 9 Applications 10 7 SERVICE MAINTENANCE DISPOSAL 11 Cleaning 11 Maintenance Service 11 Disposal 11 8 DECLARATION OF CONFORMITY 11 ...

Page 3: ...l version is not explosion protected Protect the ionizer against high air humidity temperature steams and dust Take care to select a location free of water and oil Do not expose the ionizer to strong humidity for extended periods Non permitted condensation condensation of air humidity on the appliance may occur if a cold appliance is taken to a considerably warmer environment In this case acclimat...

Page 4: ...ck see sticker on the left side Always use the genuine power pack The stated voltage value must be the same as the local voltage Risk of injury The peaks of the ion source are sharp and cutting The ionizer generates poisonous ozone ensure sufficient ventilation For maintenance and repair work disconnect the ionizer from the mains Disconnect the ionizer from the mains during periods of idleness ...

Page 5: ... in major display deviations incorrect weighing results Discharge sample at a sufficient distance from the weighing balance During normal operation the green LED will be glowing in case of a breakdown the red LED If the red LED is glowing turn off the ionizer at the master switch and turn it on again If the red LED continues to glow inform the manufacturer The ionizer may heat up slightly during o...

Page 6: ...om ion source Ion source Stainlees steel Maximum air volume 0 06 cm3 min Weight 310 g Dimensions cm 11 x 10 5 x 6 Ambient conditions 0 50 C 20 80 air humidity non condensing AC adapter primary AC 100 240V 50 60Hz Rated electric power supply DC 12V 500 mA Pollution Degree 2 Overvoltage Category Category II Altitude Up to 2000 m Installation Site device may only be used indoors ...

Page 7: ... BA e 1620 7 4 Appliance overview 1 Ion source 2 LED Ionizer operation green normal operation red breakdown 3 LED fan operation yellow on 4 Fan on and off switch 5 Mains adapter connection 6 Fan 7 Main switch ...

Page 8: ...issioning Unpacking Remove device carefully from packaging remove plastic wrapping and place device at the planned work place Scope of delivery 1 Ionizer 2 Mains adapter 3 Operating instructions Placing With adjustable feet Without adjustable feet ...

Page 9: ...D starts glowing During normal operation the green LED will be glowing in case of a breakdown the red LED If the red LED is glowing turn off the ionizer at the master switch and turn it on again If the red LED continues to glow disconnect the device from the mains and contact the manufacturer After use turn off with master switch OFF Turning on the fan Ionisation results are better when the fan is...

Page 10: ...eriod required for discharging the sample will be shortened Discharge of powdery samples Discharge prevents whirling up problems with toxic samples Turn off fan for easily whirled up samples Discharge of goods to be weighed glass wind screen or weighing containers Discharge of beakers etc Powdery samples are deposited at the inner wall of the glass when being poured into the beaker This is prevent...

Page 11: ...residues powder carefully with the help of a brush or hand held vacuum cleaner Keep ion source and fan clean keep holes placed on front part of the ionizer clean Maintenance Service The appliance may only be opened by trained service technicians who are authorized by KERN Before opening disconnect from power supply Disposal Disposal of packaging and appliance must be carried out by operator accord...