YBI-01A-BA-e-1620
9
Mains connection
Select a country-specific power plug and insert it in the
power unit.
Power is supplied via the external mains adapter. The
stated voltage value must be the same as the local
voltage.
Only use original KERN mains adapters. Using other
makes requires consent by KERN.
Turn off the ionizer (master switch “OFF“) before
connecting it to the mains adapter.
6
Commissioning
Turning on the ionizer
Set master switch to “ON“.
The green LED starts glowing.
During normal operation the green LED will be glowing, in
case of a breakdown the red LED.
If the red LED is glowing, turn off the ionizer at the master
switch and turn it on again. If the red LED continues to
glow, disconnect the device from the mains and contact
the manufacturer.
After use turn off with master switch "OFF".
Turning on the fan
Ionisation results are better when the fan is turned on and
the period required for discharging the sample will be
shortened.
Information:
Turn off fan for powdery, easily whirled up samples.
Set switch of fan to “ON”.
The yellow LED starts glowing.
Ensure sufficient ventilation even when the fan is turned
on.