background image

08/2016 Rev. 1.0 

Page 3

SiPM

Eval Kit

1 Introduction

The KETEK SiPM Evaluation Kit allows an easy operation and evaluation of any KETEK SiPM. It can be used 

for a wide range of applications which require e.g. single photon counting or measurements with scintilla

-

tors. Two of the Evaluation Kits can be mounted face to face for coincidence measurements.
The SiPM Evaluation Kit is equipped with a preamplifier and set up modularly so that the PCBs with the 

presoldered SiPMs can be swapped to a different SiPM model.
For the operation of the Evaluation Kit only a +12 V DC power supply, a bias source and an oscilloscope are 

required.

For full technical details of the SiPMs and the Evaluation Kit, please refer to the 

KETEK SiPM Datasheet and 

the KETEK SiPM Evaluation Kit Datasheet.

Additionally, KETEK provides a SiPM Bias Source which can be directly connected to the Evaluation Kit. 

Please refer to the 

KETEK SiPM Bias Source Datasheet 

for further information.

2 Setup Example

• 

 The signal and bias connectors are SMA type, the preamplifier is equipped with solder pins. An over

-

view is shown in figure 1. Signal lines need to be be terminated with 50 Ω.

• 

Preamplifier power: +12V DC, 150 mA

• 

Bias: positive with max. + 40 V, typical current limit 2 mA

• 

Signal: connected to the preamplifier 

• 

Note:

 Signal corresponds to 90% of the SiPM signal

• 

Monitor: connected to an oscilloscope (50 Ω DC)

• 

Note:

 Monitor corresponds to 10% of the SiPM signal

• 

Amplified Signal:  connected to an oscilloscope (50 Ω DC)

• 

 Since the SiPM is a highly sensitive photodetector, it must be operated under dark conditions. 

• 

 After biasing the SiPM with e.g. 4 V above the breakdown voltage, dark counts should be visible at the 

Amplified Signal path, using e.g. a timebase of 100 ns/div and a vertical resolution of 20 mV/div. The 

Monitor is used for higher photon fluxes for which the preamplifier saturates (e.g. when coupling a 

bright scintillator).

• 

 Note: The procedure may vary depending on the used readout electronics. For the examples shown here, a 

digital oscilloscope is used.

Fig. 1 SiPM Evaluation Kit Connection Scheme

Amplified 

Signal

+12 V DC

GND

Preamplifier

PCB with SiPM

Bias

Monitor

Signal

Mounting Holes for 

Coincidence Setup 

 

www.aptechnologies.co.uk

AP 

Technologies Limited   The Coach House   Watery Lane   Bath   BA2 1RL

 

T:  +44 (0) 1225 780400    F:  +44 (0) 8701 266449    E:  info@aptechnologies.co.uk 

Summary of Contents for SiPM

Page 1: ...www ketek net info ketek net Hofer Str 3 81737 Munich Germany phone 49 89 673 467 70 fax 49 89 673 467 77 KETEK GmbH www aptechnologies co uk AP Technologies Limited The Coach House Watery Lane Bath BA2 1RL T 44 0 1225 780400 F 44 0 8701 266449 E info aptechnologies co uk ...

Page 2: ...easurements 4 3 1 Dark Count Spectrum 4 3 2 Single Photon Spectrum 4 3 3 Using the Full Dynamic Range 4 4 Measurements with Scintillator 5 4 1 Energy Spectrum 5 4 2 Coincidence Time Resolution 6 5 Appendix 7 5 1 Variant for Optical Bench Mount 7 5 2 SiPM Bias Source 7 SiPM Eval Kit www aptechnologies co uk AP Technologies Limited The Coach House Watery Lane Bath BA2 1RL T 44 0 1225 780400 F 44 0 8...

Page 3: ... An over view is shown in figure 1 Signal lines need to be be terminated with 50 Ω Preamplifier power 12V DC 150 mA Bias positive with max 40 V typical current limit 2 mA Signal connected to the preamplifier Note Signal corresponds to 90 of the SiPM signal Monitor connected to an oscilloscope 50 Ω DC Note Monitor corresponds to 10 of the SiPM signal Amplified Signal connected to an oscilloscope 50...

Page 4: ...ed by the same optical pulse e g with an optical beam splitter The other path for the single photon measurement can be attenuate e g with neutral density filters Measure the area respectively the charge of the SiPM signal in e g a 25 ns wide gate The histogram of the measured charge is the single photon spectrum cf fig 2 Note E g the relative gain breakdown voltage and photo detection efficiency c...

Page 5: ...ge released by the SiPM corresponding to the deposited energy in the scintillator Evaluation Kits can be ordered either with unmounted or premounted LYSO crystals on the SiPMs For PM11 1 2 x 1 2 mm active area LYSO with 1 2 x 1 2 x 4 0 mm For PM33 3 0 x 3 0 mm active area LYSO with 3 0 x 3 0 x 5 0 mm For PM66 6 0 x 6 0 mm active area LYSO with 5 0 x 5 0 x 5 0 mm 4 1 Energy Spectrum In analogy to s...

Page 6: ...he minimum between Compton edge and 511 keV photopeak to filter for the energy respectively coincident events The Coincidence Timing can then be measured by the time difference of both Amplified Signals The output of the coincidence logic triggers the measurement of the timestamps of both Amplified Signals Note Best timinig is achieved with a timing measurement close to the baseline Typically with...

Page 7: ... the Evaluation Kit 5 2 SiPM Bias Source As an addon to both Evaluation Kits a SiPM Bias Source is available cf fig 5 It offers an adjustable bias voltage from 20 to 40 V with either positive or negative polarity Current limit can be set to either 2 or 20 mA For full technical specifications please refer to the KETEK SiPM Bias Source Datasheet Fig 5 SiPM Evaluation Kit for Optical Bench Mount with...

Reviews: