20
Display
Operating instructions
4
Performing a tare
Shut the heater lid, check to make sure that the steady-state mark ( ) is
displayed, and press the
key. A series of hyphens and the word
‘TARE’ will be displayed, the sample dish will move vertically and a tare will
be performed. A sharp tone will then sound and the word ‘TARE’ will
disappear from the display, a weight of ‘0.000 g’ will be displayed and the
tare will then be complete.
*
Always be sure that the heater lid is closed when performing a tare. Also make
sure that the tester is not exposed to drafts or breezes, or subjected to any
vibrations when a tare is being performed.
5
Placing a sample in the tester
Open the lid to the heater and place the sample inside.
Be sure to place the sample as flatly as possible into the tester so that heat
is applied evenly to the sample during measurement. (See “7. Tips on how
to ensure accurate measurements” on p. 17 for further information.)
6
Beginning measurement
Close the lid of the heater, check to make sure that the stability indicator ( )
is being displayed, and then press the
key. The display will switch
from a display of the weight (i.e., grams) to a percentage display and the
measuring time will be displayed.
The heating indicator (
) will start to flash and drying will begin.
*
There may be times when the stability indicator (
) is not displayed because of
external vibrations or drafts or breezes. While it is possible to perform
measurements at such times, there may be cases when it is impossible to
obtain accurate measurements at such times. You should always be sure to
conduct measurements in a location as free as possible of vibrations, drafts,
breezes, or any other harmful influences.