26
2) Specifying settings for TIME (Timed halting) mode
In timed halting mode, the measuring time is specified before the test and the sample is dried to determine
its moisture (or solid) content. When the specified measuring time is reached, the test is halted.
Measuring time may be specified in 1-minute increments anywhere in a range from 1 to 240 minutes or
measuring may be specified to be performed over a continuous period of 12 hours.
Key(s) used
Operating instructions
With the display displaying the weight in grams, press the
key.
If a password has been specified, the message ‘PASS’ will be
displayed to indicate that a password should be entered. Follow
the instructions in “10-2-7 Specifying a password” on p. 46 to
enter the password.
The ‘CONDITIONS’ display area will begin to flash. Press the
key until ‘MODE’ begins to flash.
With the ‘MODE’ display area flashing, press the
key. This
will cause the highlight to shift to measuring mode selection and
for the currently specified measuring mode to begin flashing.
Press the
key until ‘TIM’ begins to flash.
*
If ‘TIM’ is already flashing, then proceed directly on to the next
step.
When ‘TIM’ begins to flash, press the
key. This will cause
the measuring time settings to become active and for the currently
specified measuring time to begin to flash.
Press the
key or
key to select the desired measuring
time. Note that here a time of ‘30 minutes’ is specified.
When the desired measuring time begins flashing, press the
key. The specified measuring time will change from a
flashing to a constantly lit display and ‘MODE’ will begin to flash
instead.
If you wish to specify other measuring conditions settings, press
the
key to select the setting to be specified. To exit from
specifying measuring conditions settings, press the
key
instead.
Step
1
2
3
4
5
6
7
8
9
Display
Summary of Contents for FD-720
Page 1: ...Infrared Moisture Determination Balance FD 720 Operating Manual ...
Page 2: ......
Page 66: ......
Page 67: ......
Page 68: ...0706 PA 0201 200 ...