29
4) Specifying settings for SLOW (Low-speed drying) mode
In low-speed drying mode, the temperature is raised more gradually than when performing measurements
under normal conditions, with it taking about 5 minutes from the time when measuring begins until the
temperature reaches the specified drying temperature.
Either ‘AUTO’ (automatic halting mode) or ‘TIME’ (timed halting mode) may be specified to define the
conditions under which measuring is to be halted.
Key(s) used
Operating instructions
With the display displaying the weight in grams, press the
key.
If a password has been specified, the message ‘PASS’ will be
displayed to indicate that a password should be entered. Follow
the instructions in “10-2-7 Specifying a password” on p. 46 to
enter the password.
The ‘CONDITIONS’ display area will begin to flash. Press the
key until ‘MODE’ begins to flash.
With the ‘MODE’ display area flashing, press the
key. This
will cause the highlight to shift to measuring mode selection and
for the currently specified measuring mode to begin flashing.
Press the
key until ‘SLW’ begins to flash.
*
If ‘SLW’ is already flashing, then proceed directly on to the next
step.
When ‘SLW’ begins to flash, press the
key. The highlight
will then shift to specification of low-speed drying mode settings
where the halting conditions must be specified. The units area for
timed halting mode (‘min’) or automatic halting mode (‘%’) will
begin to flash.
Pressing the
key will cause the units to switch between
minutes (‘min’) and a percentage (‘%’). Press the
key until
the desired units are displayed.
When the desired units for the halting conditions (i.e., minutes or
a percentage) begin to flash, press the
key. This will cause
the highlight to shift to the specification of the halting conditions
and the numerical field begin to flash.
Press the
key or
key to select the desired halting
conditions. In this example, timed halting (20 minutes) is
specified.
When the desired time or percentage flashes, press the
key. The specified halting conditions will then change from
flashing to a constantly lit display and ‘MODE’ will begin to flash.
If you wish to specify other measuring conditions settings, press
the
key to select the setting to be specified.
To exit from specifying measuring conditions settings, press the
key instead.
Step
1
2
3
4
5
6
7
8
9
0
!
Display