41
10-2-3 Specifying sample codes
This section describes how to specify sample codes to be used when outputting measurements to
a printer or computer. Sample codes may be specified as 4-character codes.
Any digit from 0~9, and letter from A~Z, or a hyphen (i.e., ‘—’) may be specified as the value of
the first or second character of the code.
Only digits from 0~9 may be specified as the value of the third or fourth character of the code.
* The digits in the third and fourth characters of the code are automatically incremented after each
measurement, with the value returning to ‘00’ when it would otherwise exceed a value of ‘99’.
How to specify sample codes
Key(s) used
Operating instructions
With the display displaying the weight in grams, press the
key.
Press the
key until the ‘CODE’ menu item is displayed.
Press the
key. The highlight will then shift to the
specification of the sample code and the first character of the
currently selected ‘CODE’ will begin to flash.
Press the
key or
key to select a digit from ‘0’ to ‘9’, a
letter from ‘A’ to ‘Z’, or a hyphen. Note that here ‘A’ has been
selected.
When the desired character has been displayed, press the
key. The first character will then have been specified and the
second character will begin to flash.
Repeat steps
4
and
5
until the fourth character has been
specified.
When the fourth character has been specified, press the
key.
The sample code just specified will then be displayed and the
display will return to the state described in step
2
.
If you wish to specify other measuring conditions settings, press
the
key to select the setting to be specified.
To exit from specifying measuring conditions settings, press the
key instead.
Step
1
2
3
4
5
6
7
8
Display