2-6 Operating error
Loop impedance(IEC61557-3)
The influencing variations used for calculating the Operating
error are:
2.7 General specification
Operating
0 to 40
℃
,relative humidity
temperature and
85%or less, no condensation.
humidity.
Storage
-20 to+60
℃
,relative humidity
temperature and
85%or less, no condensation.
humidity.
Battery type
Eight AA R6 or LR6 batteries
Measurement times:
approx 1500 times or more.
(ATT 500 times or more.)
Dimensions:
175 x 115 x 85.7mm
Weight:
810g
Maximum altitude:
2000m
Over range indication:
'OL'
8
Range
Operating range compliant
with EN61557-3 operating error
20
Ω
0.35 to 19.99
Ω
200
Ω
20.0 to 199.9
Ω
2000
Ω
200 to 1999
Ω
Ambient temperature:
0
゜
and 35
℃
Phase angle:
0
゜
to 18
℃
System frequency:
49.5Hz to 50.5Hz
System voltage:
195V to 253V
Supply voltage:
8V to 13.8V