38
Keysight 34931A-34933A User’s Guide
34933T-002 Terminal Block for One-Wire Mode
This terminal block with screw-type connections is labeled with the model number
and the abbreviated module name. In addition, space is available on the label for
you to write the slot number.
NOTE
All modules that connect to the internal DMM are interlock protected. This
means that when an installed module is exposed (no terminal block or cable is
connected), the Analog Bus relays and current channels are open and
disconnected from the Analog Buses. See
for further information.
NOTE
If you are using an Keysight terminal block to connect your DUT to this module
be sure to use the 34933T-002 terminal block that corresponds to the 1-wire
configuration mode. Note that an error will not be generated if you have installed
a terminal block that doesn't match the present module configuration.
NOTE: Analog Bus
connections are
on Matrix 3 and
Matrix 4.
Wire Size:
20 AWG Typical
18 AWG Max