8163A/B, 8164A/B, 8166A/B User’s Guide
103
Power Measurement
3
To set the calibration offset to 40.000 dB:
1 Move to the Power Meter channel and press the [Details] softkey.
2 Move to the [
Cal
] parameter and press
Enter.
3 Enter 40.000 and press
Enter
.
How to Set the Reference Level
How to Input a Reference Level
dB results are shown relative to a reference level. The [
Ref
] parameter sets
the reference level. Setting, or changing, the reference only affects results
that are displayed in dB.
The reference level is displayed in units of dBm or Watts. Depending on
which is selected the following equations are used to calculate the power
level in dB:
P
display
(dB) = P
measured
(dBm)
-
REF(dBm)
or
Where,
P
display
is the displayed relative power,
P
measured
is the absolute power level (see
How to Set the Calibration Offset
REF
is the reference level.
You can choose the units for the reference using the Power Unit softkey. If
you choose dB, the power value changes to dB and the displayed
reference level changes to dBm.
To set the reference level to 10 mW:
1 Move to the Power Meter channel and press the [Details] softkey.
2 Move to the [
Ref
] parameter and press the [Power Unit] softkey.
3 Move to <
W
> and press
Enter
twice.
4 Change the units to mW using the [Unit+] and [Unit
-]
softkeys.
5 Enter 10.000 and press
Enter
.
P
display
dB
(
)
10
P
measured
W
( )
REF W
( )
-----------------------------------
log
=
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......