104
8163A/B, 8164A/B, 8166A/B User’s Guide
3
Power Measurement
How to Set the Reference Value to the Current Power Value
In addition to entering a new reference value, you can change the
reference value to the currently displayed power value by pressing the
[Dsp->Ref] softkey.
Pressing the [Dsp->Ref] softkey takes the input power level and stores it as
the reference. Setting the reference only affects results displayed in dB.
When you press the [Dsp->Ref] softkey, the absolute power level, the
power value in dBm or Watts, is stored as the reference, that is:
REF = P
measured
Where,
REF is the reference, and
P
measured
is the absolute power level (see
How to Set the Calibration Offset
The [Dsp->Ref] softkey is displayed if you move to a power measurement
channel.
How to Reference Another Power Measurement Channel
You can use the reference mode, [
Ref Mode
], to reference another Power
Meter’s current power level. If you choose this mode, the reference of the
channel continuously updates to the current power value measured by the
other channel.
The following equations describe how the power in dB is calculated if you
reference another channel:
P
display
(dB) = P
measuredA
(dBm)
-
P
measuredB
(dBm)
-
REF(dB)
or
Where:
P
display
is the displayed relative power,
P
measuredA
is the absolute power level (see
on page 102) measured by the current Power Meter, and
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......