110
8163A/B, 8164A/B, 8166A/B User’s Guide
3
Power Measurement
How to Set the Range
If you choose <
Auto
> from the Range mode menu, this parameter can not
be set. The Range parameter, [
Range
], is displayed in light grey text, see
on page -110, in automatic ranging mode and displays the
automatically-determined range.
Figure 62
Auto-Range Mode
If you choose <
Manual
> from the Range mode menu, you must choose a
range setting from the Range menu.
To choose a range setting, follow the following steps:
1 Setup the instrument.
2 Move to the Power Meter channel and press the [Details] softkey.
3 Move to the [
P
] parameter and press the [Power Unit] softkey. You will
see the menu in Figure 57.
4 Move to <
dBm
>, by using the cursor key, and press
Enter
.
5 Move to the [
Range Mode
] parameter and press
Enter
.
6 Move to <
Auto
>, by using the cursor key, and press
Enter
.
7 Perform a set of measurements.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......