112
8163A/B, 8164A/B, 8166A/B User’s Guide
3
Power Measurement
Figure 64
Out of Range - Power Greater Than Upper Power Limit
If the measured power is more than 40 dBm less than the range setting, it
is impossible for power to be displayed. The power value, -1.--- dBm, as
shown in
on page -113, is shown. This means that the measured
power is greater than the resolution at this [
Range
] value. You must
increase the [
Range
] value.