114
8163A/B, 8164A/B, 8166A/B User’s Guide
3
Power Measurement
Figure 66
Range Value Menu
Upper Power Limit and Resolution
on page -114 shows the upper power limit and measurement
resolution at various power ranges. As can be seen the resolution
decreases as the chosen [
Range
] decreases. The resolution is always
40 dBm less than the chosen [
Range
] value. The Upper Power Limit is
always 3 dBm greater than the chosen [
Range
] value.
Table 11
Upper Power Limits and Resolution for Various Power Ranges
[R
ANGE
]
Upper Power Limit
Resolution
40 dBm
19.999 W
43.000 dBm
0.001 W
-
0.000 dBm
30 dBm
1.9999 W
33.000 dBm
0.1 mW
-
10.000 dBm
20 dBm
199.99 mW
23.000 dBm
0.01 mW
-
20.000 dBm
10 dBm
19.999 mW
13.000 dBm
0.001 mW
-
30.000 dBm