116
8163A/B, 8164A/B, 8166A/B User’s Guide
3
Power Measurement
With modules designed for the 8153A Lightwave Multimeter, for averaging
times of more than 1 second, the displayed power is given by the formula:
Where,
P
new
is the new displayed result,
P
old
is the previously displayed result,
ample
is the value read by the hardware,
T
avg
is the averaging time, as set by the user, and
T
sample
is the time the hardware takes to make a reading.
If the measurement condition change (for example, a range change in
automatic ranging),
P
old
is reset and the averaging starts again. This is why
the display update seems faster in automatic ranging.
To set the averaging time to 1 second:
1 Move to the Power Meter channel and press the [Details] softkey.
2 Move to the [
AvgTime
] parameter and press
Enter
.
3 Move to <
1 s
>, by using the cursor key, and press
Enter
.
How to Choose the MinMax Mode
MinMax mode measures the incoming power and displays the minimum
value measured, [
Pmin
], and the maximum value measured, [
Pmax
]. The
difference between these values,
D
P, is displayed in place of P, the power
value. This mode is intended principally for polarization dependent
measurements, but can be used for other types of measurement.
NOTE
For a Dual Power Sensor, you cannot set the averaging time of the slave
channel, channel 2, directly. By default, the averaging time of the slave
channel, channel 2, is the same value as that for the master channel,
channel 1. See
on page -122 for more details.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......