164
8163A/B, 8164A/B, 8166A/B User’s Guide
5
Tunable Lasers
Figure 93
Input Trigger Mode
4 Move to one of the following, by using the cursor key:
•
<Ignore
>, input triggers are ignored.
•
<
Next Step
>, an input trigger will cause the next step of a stepped
sweep to be performed.
•
<
Run Sweep
>, an input trigger will start a single sweep cycle.
5 Press
Enter.
You can generate input triggers in any of the following ways:
•
applying a trigger to the Input Trigger Connector on the rear panel
of your instrument,
•
setting <
Trigger Configuration
> to <
Loopback
> so that an output
trigger automatically generates an input trigger, or
•
using the
:TRIGger
GPIB command, see your mainframe’s
Programming Guide.