Return Loss Measurement
7
Calibration Measurements
Before measuring the reflection factor of a device under test (DUT), it is
important to take calibration measurements as described in
on page 188. These calibrations eliminate
wavelength dependencies, coupler directivity, insertion losses,
backscattering and other non-ideal characteristics of the system.
The Return Loss Application described later in the manual, available on the
8163B and 8163A, provides a convenient guide through the calibration
and measurement process.