8163A/B, 8164A/B, 8166A/B User’s Guide
189
Return Loss Measurement
7
Calibration using the Keysight 81000BR Reference Reflector
You can use the Keysight 81000BR Reference Reflector to make a
Reflectance Calibration, and a Termination Calibration.
Reflectance Calibration
Use this procedure to calibrate the Return Loss module against a
component of known return loss.
1 Making sure all the connectors are clean, set up the instrument as
shown in
on page -189 if you are using an External Source.
Figure 107
Reflectance Calibration - External Source
on page -189 if you are using an internal source.
Figure 108
Reflection Calibration - Internal Source
Tip:
For best results and higher repeatability, fix the cable.
2 Attach a component with a known return loss to the end of the
patchcord.
Laser
Keysight
81000BR
Keysight 81000FI
81000UM
8161x
8163B Lightwave Multimeter
Laser
Source
81113PC
patchcord with FC/PC
output connector
8161x
8163B Lightwave Multimeter
Keysight
81000BR
Keysight 81000FI
81000UM
8161x
8163B Lightwave Multimeter
patchcord with FC/PC
output connector