8163A/B, 8164A/B, 8166A/B User’s Guide
195
Return Loss Measurement
7
7 Press [Close] to exit from the menu.
Measuring the Power Transmitted through the Measurement Patchcord
Next, measure the power transmitted through the measurement patch
cord.
1 Making sure all the connectors are clean, set up the instrument as
shown in
on page -195 if you are using an External Source,
Figure 116
Power Transmitted through the Measurement Patchcord - External Source
on page -195 if you are using an internal source.
Figure 117
Power Transmitted through the Measurement Patchcord - Internal Source
2 .Move to the Power Sensor channel:
•
Press [Menu].
•
If necessary, move to <
Pwr unit
>, move to <
dB
> and press
Enter
.
•
The Power Sensor channel displays a power value in dB that is equal
to the front panel delta of the measurement system.
Power
Sensor
8163B Lightwave Multimeter
measurement patchcord
8161x
8163B Lightwave Multimeter
Laser
Source
Sensor
8161x
Power
8163B Lightwave Multimeter
measurement patchcord
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......