Return Loss Measurement
7
2 Terminate the measurement patchcord so that there are no reflections
coming from the end.
Tip:
You can do this by wrapping the fiber five times around the shaft of a
screwdriver (or some similar object with a diameter of between 5 mm and
7 mm).
3 Move to [
Terminated calibration
] and press
Enter
. The instrument
measures the power reflected by the component, and sets the [
Para
]
values used by the Return Loss monitor’s power sensor and monitor
diode.