200
8163A/B, 8164A/B, 8166A/B User’s Guide
7
Return Loss Measurement
Figure 122
The Calibration Parameters Screens - Return Loss Diode
2 Press [Next]. The
Calibration parameters
screen for the
Monitor Diode
appears, as displayed in
on page -201. This screen shows
current value for the following quantities:
•
[
Para
], the parasitic power value measured by the Return Loss
module’s monitor diode in dBm. This value is determined by the
termination calibraion, or you can use the default setting held in the
factory calibration.
•
[
Ref
], the power measured by the Return Loss module’s monitor
diode during the reflectance calibration in dBm
•
[
Meas
], the power currently measured by the Return loss Module’s
internal monitor diode in dBm