8163A/B, 8164A/B, 8166A/B User’s Guide
201
Return Loss Measurement
7
Figure 123
The Calibration Parameters Screens - Monitor Diode
3 Press [Next]. The
Calibration parameters
screen for
User Data
appears,
as displayed in
on page -202. This screen shows current
value for the following quantities:
•
[
FPDelta
], the loss correction, in dB, due to differences between the
reference cable and the measurement cable, such as differences in
the front panel connector.
•
[
RLref
] the value of the return loss of the reference relector used. If
you areusing a reference cable, this value will be around 14.6 dB