8163A/B, 8164A/B, 8166A/B User’s Guide
205
Return Loss Measurement
7
Measuring the Power when there are No Reflections
Next, terminate the measurement patchcord so that there are no
reflections from the end. All the power measured by the sensor now, is due
to the non-ideal nature of the measurement system. This is our
termination parameter.
Figure 127
Measuring the Power with the Connector Terminated
This measured power for the termination parameter is called
P
T
.
Measuring the Power Transmitted Through the Measurement Patchcord
Connect the measurement patch cord (the cable you will use to connect to
the Device Under Test, DUT) directly to a Power Meter and measure the
transmitted power,
E
Meas
. You can use
E
Meas
to calculate the front panel
delta, see
Calculating the Front Panel Delta
Calculating the Insertion Loss of the DUT
Figure 128
Measuring the Power transmitted through the Measurement Patchcord
Reflection-free
Termination
= 0
M
ref
P
ref
Keysight 81610A
Reflectivity
P
in
measurement patchcord
Power Meter
E
Meas
M
ref
P
ref
Keysight 81610A
Reflection Reference
measurement patchcord
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......