8163A/B, 8164A/B, 8166A/B User’s Guide
209
Return Loss Measurement
7
If we divide equation 8 by equation 7, this gives us the following equations:
Thus we can use the equation below to calculate return loss:
The return loss of the reference reflection is given by:
Calculating the Front Panel Delta
The Front Panel Delta is the change in loss variation that is caused by
replacing the reference cable, as used in
page 192, with the measurement patchcord, as used in
on page 198. This is caused by differences in reflections from
the front panel connector and also differences in the backscatter level of
the fibers.
To measure the front panel delta you must measure the power transmitted
through the reference cable,
Measuring the Power Transmitted Through
on page 204, and the power transmitted through
the measurement cable, see
Measuring the Power Transmitted Through
on page 205. The system may be represented
Summary of Contents for 8163A