8163A/B, 8164A/B, 8166A/B User’s Guide
223
Setting Attenuation and/or Power Levels
8
Figure 137
81570A/71A/73A Attenuation Offset
α
Offset
of 1.651 dB ‘added in’ to the
attenuation factor
α
to calibrate for intrinsic insertion losses in the test
setup.
81576A/77A modules
The [PM -> Off ] softkey offered by the
α
Offset
parameter is functionally
equivalent to the [PM -> Off ] softkey offered by the P
Offset
parameter. You
can use the [PM -> Off ] softkey to offset the difference between the power
measured by the internal powermeter and your reference powermeter, as
described
Calibrating Test Setups that use an 81570A/71A/73A/78A
• When the
α
Offset
parameter is selected, [PM -> Off ] writes to
α
Offset
• When the P
Offset
parameter is selected, [PM -> Off ] writes to P
Offset
These functions are mathematically identical.
Neither function is a measurement of total intrinsic insertion loss as
insertion loss of the Attenuator module itself is not included.