274
8163A/B, 8164A/B, 8166A/B User’s Guide
10
Applications
The Stability Application
The Stability data acquisition application logs a series of power
measurements for a number of Power Meter channels, plots the results as
a graph, and generates a statistical analysis of the results.
Figure 169
Example Stability Application
The Stability application differs from the
page 261 because you set the averaging time, the total time, and the
maximum number of measurements and all power measurements and the
instrument calculates a period time, a new measurement is started after
the completion of every period time. Period time may be greater than or
equal to the averaging time you set and is always greater than or equal to
100 ms.
This leaves the following two possibilities:
• The averaging time is less than the period time, there is a dwell time
between the commencement of each power measurement, as shown in
on page -274. This can happen because:
•
the maximum number of power measurements multiplied by the
averaging time is less than the total time or
•
the averaging time is less than 100 ms.
• The averaging time is equal to the period time, all power measurement
are performed without any dwell time.
Averaging Time
Period Time
t
1
2
3
4
5
Dwell Time
Total Time
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......