8163A/B, 8164A/B, 8166A/B User’s Guide
291
Applications
10
•
<
High Power (2)
>, a high power optical output, for example, for use
in tesing Erbium Doped Fiber Amplifiers (EDFAs), you see a screen
similar to
•
<
Low SSE (1)
>, a lower power optical output with low Source
Spontaneous Emission (SSE), for example, for use in passive
component test, you see a screen similar to
•
<
Both (master:2)
>, where both optical outputs can be enabled but
only the output of the high power optical output can be regulated,
and
•
<
Both (master:1)
>, where both optical outputs can be enabled but
only the output of the low SSE optical output can be regulated.
•
Press
Enter
.
How to Measure the Reference
18 Press [Reference].
The PACT Reference Screen, as shown in
appears. The number of samples that each Power Meter will record and
the output laser power are displayed If <
Continuous
> was selected as the
sweep [
Method
], the sweep speed, [
V Sweep
], is displayed.
NOTE
If [
Step
] does not divide exactly into the difference between [
λ
Stop
]
and [
λ
Start
], the parameters are automatically corrected to meet the
constraints of PACT.