8163A/B, 8164A/B, 8166A/B User’s Guide
377
Error Messages
14
Errors appearing on pop-up menus
Error on Module
error message
Calibration data corrupted
description
The calibration data loaded on the module is invalid. Either the caldata version number is not compatible
with the firmware or the caldata is corrupt (checksum incorrect)
affected modules
all modules
action
Please return the unit to your local service center for repair
error message
Laser safety error
description
The laser control circuit has switched of the laser. Either the laser current or the laser power was too high.
affected modules
TLS
action
Try to run the automatic realign from the menu or to operate the laser at a lower power level. If the failure
persists, please return the unit to your local service center for repair.
error message
Temperature out of range
description
The temperature to operate the tunable laser source module was measured outside given limits.
affected modules
TLS
action
Verify the fan is working and the ventilation system is not blocked. If the failure persists, please return the
unit to your local service center for repair.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......