error message
Zeroing failed. Light?
description
Zeroing has still detected some light. Either the input is not darkened properly or there is some error on the
powermeter module/head.
affected modules
Powermeter
action
Darken the optical input(s) properly. If the failure persists, please return the unit to your local service center
for repair.
error message
Selftest failed
description
The selftest of the module in slot <slot> has failed. If the slot is occupied by an optical head interface
module, the head selftest has failed.
affected modules
all
action
Please return the unit to your local service center for repair.
error message
BRAM corrupted
description
The battery buffered RAM is corrupted.
affected modules
Mainframe
action
Please return the unit to your local service center for repair.
error message
Please run realign (from menu)
description
The alignment of laser cavity is not ideal.
affected modules
TLS
action
Please follow this instruction and run the realignment, which can be selected from the menu of the tunable
laser source module. If the failure persists, please return the unit to your local service center for repair.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......