8163A/B, 8164A/B, 8166A/B User’s Guide
379
Error Messages
14
error message
Out of specification
description
The unit has been found outside the specification.This may happen due to various reasons.
affected modules
TLS
action
Please run LambdaZero from the menu first. If the failure persists, reboot the mainframe and run realign
from the menu. If this does not help, please return the unit to your local service center for repair.
error message
Duty cycle out of range
description
The power control loop cannot control the correct output power anymore, because the duty cycle of the
external modulation is too small.
affected modules
TLS
action
Increase the duty cycle (that is, increase the on-state duration of the modulated signal)
error message
Beampath safety error
description
A problem on the module has lead to an unsafe operation state. For this reason, the shutter has been
closed.
affected modules
All attenuators
action
If this problem occurs several times, please send the module to the local Keysight service center for repair.
error message
Module communication error
description
There was an error in communication between module and mainframe. For safety reasons, the module was
disabled and the shutter closed.
affected modules
All attenuators
action
If this problem occurs several times, please send the module to the local Keysight service center for repair.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......