8163A/B, 8164A/B, 8166A/B User’s Guide
383
Error Messages
14
error message
Initialization Error (Calibration Data Invalid, error value)
description
The calibration data was tagged as invalid before it was loaded on to the module
error value
0
not significant
action
Please return the unit to your local service center for repair
error message
Initialization Error (CalOptZeroMeas, error value)
description
Error in optical zero measurement
error value
1
not significant
action
Please return the unit to your local service center for repair
error message
Initialization Error (DspMainCalcKFactor failed, error value)
description
DSP K factor calibration failed
error value
-1
no TCFS (according to caldata)
1
error when calibrating TCFS
2
DSP sweep failed
4
laser safety
action
Please return the unit to your local service center for repair
error message
Initialization Error (DspMainWriteCaldata(), error value)
description
Error while writing cal data to DSP
error value
5
tried to access invalid array position
6
wrong dimension
7
unknown type found in variable array
8
wrong parameter when trying to access one of the DSPs cal
arrays
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......