8163A/B, 8164A/B, 8166A/B User’s Guide
385
Error Messages
14
error message
Initialization Error (MainEEPROMInit, error value)
description
Error reading main EEPROM
error value
1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (MotorSelfCali failed, error value)
description
Motor self calibration failed
error value
-1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (MotorZero, error value)
description
DSP motor zero not finished in time
error value
1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (OptZero, error value)
description
Optical zero disabled in caldata
error value
-1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (OptZero Opc, error value)
description
Timeout while waiting for laser diode (16) and thermopositioner (64) to settle while preparing
optical zero measurement
error value
16
waiting for laser diode
64
waiting for thermopositioner
action
Reboot the instrument. If the failure persists, please return the unit to your local service center
for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......