388
8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
error message
Initialization Error (Tcfs temperature not settled, error value)
description
Temperature has not settled at the tuning control feedback system
error value
0
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (TmpCalDataInit, error value)
description
Error accessing caldata
error value
1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error (UpPiezoEnable, error value)
description
Over current
error value
1
(not significant)
action
Please return the unit to your local service center for repair
error message
Initialization Error - module temperature out of range
(please wait a little bit and try again)!
description
Module temperature is too high
error value
(no error value)
action
Doublecheck whether the fan is working and not blocked. Wait some time and reboot the
instrument. If the failure persists, please return the unit to your local service center for repair.