390
8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
error message
TCFS Temp. Selftest Failed (error value)
description
Verifies if the peltier temperature is within target with a given uncertainty
error value
15010
TCFS peltier temperature is not in target range
action
Please return the unit to your local service center for repair
error message
TCFS Resistor Selftest Failed (error value)
description
This test measures the R
d
of the photodiodes at channel A, B and M
error value
15021
Photodiode R
d
in channel A is below limit
15022
Photodiode R
d
in channel B is below limit
15023
Photodiode R
d
in channel C is below limit
action
Please return the unit to your local service center for repair
error message
TCFS Attenuator Selftest Failed (error value)
description
Verifies whether an attenuator position is repeatable.
error value
15030
Attenuator could not be positioned
action
Please return the unit to your local service center for repair
error message
Main DC Selftest Failed (error value)
description
Tests all fixed DC voltages available at the mainboard mux
error value
23011
+15V Measurement DC voltage not in limits
23012
-15V Measurement DC voltage not in limits
23013
+5V Reference DC voltage not in limits
23014
-5V Reference DC voltage not in limits
23015
negative supply voltage of laser current source not in limits
23016
positive pezo supply voltage not in limits
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......