392
8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
error value
23071
test laser current not in limits
23072
current laser safety did not trigger at overcurrent
action
Please return the unit to your local service center for repair
error message
Main Piezo Amp. Selftest Failed (error value)
description
Verifies if the piezo amplifier reaches the upper and lower limit. An error may indicate a defect amplifier
transistor
error value
23091
piezo amplifier could not set to maximum voltage
23092
piezo amplifier could not set to minimum voltage
action
Please return the unit to your local service center for repair
error message
Main Cavity Motor Selftest Failed (error value)
description
Verifies the functionality of the cavity motor
error value
23081
indicates an open motor connector
23082
indicates that a motor coil is defect
23083
indicates a short in a motor coil
action
Please return the unit to your local service center for repair
error message
TCFS DC Selftest Failed (error value)
description
Tests all fixed DC voltages available at the TCFS board mux
error value
25011
+10V measurement voltage on TCFS board not in limits
25012
-10V measurement voltage on TCFS board not in limits
25013
+5V reference voltage on TCFS board not in limits
25014
-5V reference voltage on TCFS board not in limits
25015
+2.5V reference voltage on TCFS board not in limits
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......