8163A/B, 8164A/B, 8166A/B User’s Guide
393
Error Messages
14
25016
+5V attenuator encoder voltage on TCFS board not in limits
25017
+2.5V attenuator encoder voltage on TCFS board not in limits
action
Please return the unit to your local service center for repair
error message
TCFS Temperature Selftest Failed (error value)
description
Verifies whether the TCFS temperature sensors are working correctly
error value
25021
TCFS board temperature sensor U620 above limit
25022
TCFS case temperature sensor possibly not connected
action
Please return the unit to your local service center for repair
error message
TCFS Amp. Offset Selftest Failed (error value)
description
Checks whether an applied offset for channel A, B and M is measured within limits
error value
25031
TCFS Chan A offset not in limits
25032
TCFS Chan B offset not in limits
25034
TCFS Chan M offset not in limits
action
Please return the unit to your local service center for repair
error message
TCFS Amplifier Selftest Failed (error value)
description
This test calibrates the TCFS A, B and M channel and verifies if all resulting values are in limits
error value
25041
TCFS Hybrid amplifier offset could not be removed
25042
One of the channels A, B or M failed during calibration
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......