396
8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
error message
E1x031: 0.25 ref out of limits (a, b, c)
description
The 0.25V reference voltage measured through the 16bit ADC is out of the acceptable
limits
test value
a
shows the ADC reading (0.25V = 3276)
b
shows the lower limit
c
shows the upper limit
x
refers to the internal PM channel tested
action
Please return the unit to your local service center for repair
error message
E10040: 10bit ADC test failed (a, b, c)
description
The 10bit ADC has returned a value other than 512
test value
a
shows the ADC reading
b
shows the lower limit
c
shows the upper limit
action
Please return the unit to your local service center for repair
error message
E10051: Laser <a> temp out of limits (b)
description
The laser temperature is outside the allowed range
test value
a
laser number
b
measured laser temperature
action
Verify whether the module is operated within the specified ambient temperature. If the
failure persists, please return the unit to your local service center for repair
Summary of Contents for 8163A