8163A/B, 8164A/B, 8166A/B User’s Guide
399
Error Messages
14
error message
E10080: Laser <a> control for safety test failed
description
To test the current safety comparator the laser current needs to be set to two different
values
test value
a
laser number
action
Please return the unit to your local service center for repair
error message
E10081: Safety comparator <a> test failed (b, c)
description
Checks whether laser current safety comparator includes at the right conditions a safety
interrupt at the CPU
test value
a
laser number
b
shows whether the interrupt flag was raised when the current was lower than
the limit
c
shows whether the interrupt flag was raised when the current was higher
than the limit
action
Please return the unit to your local service center for repair