400
8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
Power Meters & Interface Modules with Optical Head
error message
E10010: Invalid Caldata
description
The Caldata loaded on the module is invalid. Either the caldata version number is not
compatible with the firmware or the caldata is corrupt (checksum incorrect)
test value
(no test value)
action
Please return the unit to your local service center for repair
error message
E10012: Dual ported RAM write/read test failed
description
Writing bit patterns and reading them back to/from the dual ported ram failed
test value
(no test value)
action
Please return the unit to your local service center for repair
error message
E10013: FPGA not correctly initialized
description
Loading the FPGA with data failed. This indicates a defective FPGA or an FPGA access
problem
test value
(no test value)
action
Please return the unit to your local service center for repair
error message
E1x020: Communication with head failed
description
An external head is available but it was not possible to access the registers of the
head’s FPGA
test value
x
refers to the module channel tested
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......