8163A/B, 8164A/B, 8166A/B User’s Guide
403
Error Messages
14
error message
E1x052: TEC comparator test failed (a, b, c, d)
description
Checks the TEC regulation voltage comparator
test value
a
shows the 10bit ADC reading for the high output voltage
b
shows the 10bit ADC reading for the low output voltage
c
shows the lower limit
d
shows the upper limit
x
refers to the module channel tested
action
Please return the unit to your local service center for repair
error message
E1x053: TEC default temp invalid (a)
description
Verifies whether the default temperature is within a given range
test value
a
shows the default temperature
x
refers to the module channel tested
action
Please return the unit to your local service center for repair
error message
E1x054: TEC control failed (a, b, c)
description
When the TEC is enabled the temperature must move towards the default temperature,
coming from a previous temperature before the TEC was enabled. The moving direction
must be correct and the overshoot must be less than a certain limit
test value
a
shows the temperature before the TEC was enabled
b
shows the actual temperature while TEC is enabled
c
shows the target temperature, where b should move to
x
refers to the module channel tested
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......