406
8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
b
shows the time when the temperature regulation has reached the default
temperature. Waits maximum 30 seconds
x
refers to the module channel tested
action
Please return the unit to your local service center for repair
error message
E1x081: TEC current to temperature mismatch
description
If the TEC temperature difference does not correspond to the measured Peltier current
this test will fail
test value
x
refers to the module channel tested
action
Verify whether the module is operated within the specified ambient temperature. If the
failure persists, please return the unit to your local service center for repair.