8163A/B, 8164A/B, 8166A/B User’s Guide
409
Error Messages
14
error message
E10051: Laser <a> temp out of limits (b)
description
The laser temperature is outside the allowed range
test value
a
laser number
b
measured laser temperature
action
Verify whether the module is operated within the specified ambient temperature. If the
failure persists, please return the unit to your local service center for repair
error message
E10060: Laser <a> control failed (b, c, d, e)
description
Checks the laser current source
test value
a
laser number
b
measured current at low power and normal current source setting
c
measured current at low power and high current source setting
d
measured current at high power and normal current source setting
e
measured current at high power and high current source setting
action
Please return the unit to your local service center for repair
error message
E10070: Peltier <a> temp diff amp failed
description
Checks whether the comparator output changes from low to high values while the
temperature DAC is set to two different temperatures
test value
a
laser number
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......