410
8163A/B, 8164A/B, 8166A/B User’s Guide
14
Error Messages
error message
E10071: Peltier <a> current control test failed (b, c, d, e)
description
The Peltier temperature is set close to the ambient temperature. Using the current
offset circuit, two different currents are set and measured. If the differences between
vary more than a specific percentage this test fails
test value
a
laser number
b
delta current setting
c
delta current measured
d
actual difference between b and c
e
allowed difference between b and c
action
Please return the unit to your local service center for repair
error message
E10080: Laser <a> control for safety test failed
description
To test the current safety comparator the laser current needs to be set to two different
values
test value
a
laser number
action
Please return the unit to your local service center for repair
error message
E10081: Safety comparator <a> test failed (b, c)
description
Checks whether laser current safety comparator includes at the right conditions a
safety interrupt at the CPU
test value
a
laser number
b
shows whether the interrupt flag was raised when the current was lower
than the limit
c
shows whether the interrupt flag was raised when the current was higher
than the limit
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......