8163A/B, 8164A/B, 8166A/B User’s Guide
411
Error Messages
14
error message
E10090: Peltier <a> regulation failed (b, c, d)
description
During boot time, when the Peltier control was disabled it is enabled for two seconds
and the temperature moving direction is checked. If the movement delta is not
approaching the target default temperature or if the temperature has not reached the
default within the given ranges this test fails
test value
a
laser number
b
default temperature
c
temperature with disabled peltier
d
temperature with enabled peltier
action
Please return the unit to your local service center for repair
error message
E10091: Peltier <a> regulation failed (b, c)
description
Verifies whether the temperature is around the default value with the given limits
test value
a
laser number
b
default temperature
c
actual temperature
action
Please return the unit to your local service center for repair
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......