8163A/B, 8164A/B, 8166A/B User’s Guide
427
Cleaning Instructions
15
What do I need for proper cleaning?
on page 427 is necessary for cleaning
your instrument. For certain cleaning procedures, you may also require
certain
Standard Cleaning Equipment
Before you can start your cleaning procedure you need the following
standard equipment:
•
•
•
•
•
•
Dust and shutter caps
All of Keysight Technologies’ lightwave instruments are delivered with
either laser shutter caps or dust caps on the lightwave adapter. Any cables
come with covers to protect the cable ends from damage or
contamination.
We suggest these protected coverings should be kept on the equipment at
all times, except when your optical device is in use. Be careful when
replacing dust caps after use. Do not press the bottom of the cap onto the
fiber too hard, as any dust in the cap can scratch or pollute your fiber
surface.
If you need further dust caps, please contact your nearest Keysight
Technologies sales office.
Isopropyl alcohol
This solvent is usually available from any local pharmaceutical supplier or
chemist's shop.
If you use isopropyl alcohol to clean your optical device, do not
immediately dry the surface with compressed air (except when you are
cleaning very sensitive optical devices). This is because the dust and the
dirt is solved and will leave behind filmy deposits after the alcohol is
evaporated. You should therefore first remove the alcohol and the dust
with a soft tissue, and then use compressed air to blow away any
remaining filaments.
Summary of Contents for 8163A
Page 17: ......
Page 86: ...8163A B 8164A B 8166A B User s Guide 85 Additional Information 2 Figure 47 MAC Address...
Page 99: ......
Page 133: ......
Page 213: ......
Page 298: ...8163A B 8164A B 8166A B User s Guide 297 Applications 10 Figure 183 Low Dynamic Range...
Page 310: ...8163A B 8164A B 8166A B User s Guide 309 Applications 10 Figure 190 Printed Results...
Page 313: ......
Page 345: ......
Page 453: ......