SYSTem Subsystem
The SYSTem subsystem controls instrument functions that are not directly related to output control,
measurement, or status functions. Note that IEEE-488 Common commands also control system
functions such as state management.
SYSTem:BEEPer[:IMMediate]
Issues a single beep.
Parameter
Typical Return
(none)
(none)
Issue a single beep: SYST:BEEP
This command overrides the current beeper state . You can issue a beep even if the beeper is turned
off by
.
Programmed beeps may be useful for program development and troubleshooting.
SYSTem:BEEPer:KCLick ON|1|OFF|0
SYSTem:BEEPer:KCLick?
Disables or enables the click tone heard when you press a front-panel key.
Parameter
Typical Return
ON|1|OFF|0
0 (OFF) or 1 (ON)
Disable front-panel key click: SYST:BEEP:KCLick OFF
The front-panel key click and instrument beeper settings do not affect each other.
This parameter is not affected by
or
.
SYSTem:BEEPer:STATe ON|1|OFF|0
SYSTem:BEEPer:STATe?
Disables or enables the beep heard when an error is generated.
Parameter
Typical Return
ON|1|OFF|0
0 (OFF) or 1 (ON)
Disable the beeper state: SYST:BEEP:STAT OFF
The front-panel key click and instrument beeper settings do not affect each other.
A beep is always emitted (even with beep state OFF) when
is sent.
This parameter is not affected by
or
.
Keysight AC6800B Series Operating and Service Guide
151
3 SCPI Programming Reference
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...