[SOURce:]VOLTage:PROGramming:SOURce INTernal|EXTAC|EXTDC
[SOURce:]VOLTage:PROGramming:SOURce?
Sets the voltage programming source.
Parameter
Typical Return
INTernal|EXTAC|EXTDC
INT, EXTAC, or EXTDC
Set the voltage programming source to external AC:
VOLT:PROG:SOUR EXTAC
The EXTAC and EXTDC modes require the analog interface card.
The
command sets this parameter to INTernal.
[SOURce:]VOLTage:RANGe[:UPPer] 155|310|MINimum|MAXimum
[SOURce:]VOLTage:RANGe[:UPPer]? [MINimum|MAXimum]
Sets the voltage range. If the voltage range is switched, the
and
settings
are cleared, and ABORt[:ALL] is applied.
Parameter
Typical Return
155|310|MINimum|M2.02
Set the voltage range to 310 V
VOLT:RANG MAX
You may only set this parameter if output is OFF.
Setting this parameter turns voltage autoranging OFF and aborts both TRANsient and ACQuire
operations.
The
command sets this parameter to 155.
If you enter a value between 0 and 155, the 155 V range will be chosen. If you enter a value above 155
V, the 310 V range will be chosen.
[SOURce:]VOLTage:RANGe:AUTO ON|1|OFF|0
[SOURce:]VOLTage:RANGe:AUTO?
Enables or disables voltage autoranging. If this parameter changes, the
and
settings are cleared, and all operations are aborted.
Parameter
Typical Return
ON|1|OFF|0
0 (OFF) or 1 (ON)
Turn on voltage autoranging:
VOLT:RANG:AUTO 1
The
command sets this parameter to OFF.
3 SCPI Programming Reference
164
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...